RATINGS
Voltage and Current: 80 volts maximum, 80 amps maximum.
Power Dissipation: 320 watts maximum continuous at 28oC. Derating to 300 watts
continuous at 40oC.
Min. Operating Volts: Increases with current, <0.5V at 10A, <1V at 40A,
<2V at 80A.
OPERATING MODES
Constant Current Mode
Setting Range: 0 to 80A in 2 ranges, set by 10 turn control.
Meter Resolution: 1mA on 8A range, 10mA on 80A range.
Setting Accuracy: ? (0.2% + 20mA).
Regulation: <10mA for a 20W to 280W power change.
Slew Rate Range: <8mA/ms to >2500A/ms
Constant Resistance Mode
Setting Range: 0.04 Ohms to 100 Ohms in 2 ranges, set by 10
turn control.
Meter Resolution: 10m Ohms on 10 Ohms range, 100m Ohms on 400 Ohms range.
Setting Accuracy: ? (0.5% + 2 digits).
Regulation: <2% for a 20 Ohms to 280 Ohms power change.
Slew Rate Range: <3.3 Ohms/s to >4000 Ohms/ms
Constant Conductance Mode
Setting Range: 0.01A/V to 40A/V in 2 ranges, set by 10 turn
control.
Meter Resolution: 0.001A/V or 0.01A/V depending on range.
Setting Accuracy: ? (0.5% + 2 digits).
Regulation: <2% for a 20W to 280W power change.
Slew Rate Range: <33mA/V/s to >400A/V/ms
Constant Voltage Mode
Setting Range: Vmin to 80V in 2 ranges, set by 10 turn
control. Vmin varies with current - see Ratings.
Meter Resolution: 1mV on 8V range, 10mV on 80V range.
Setting Accuracy: ? (0.2% + 2 digits).
Regulation: <10mV for a 20W to 280W power change.
Slew Rate Range: <2.5V/s to >800V/ms
Constant Power Mode
Setting Range: 0 to 320W, set by 10 turn control.
Meter Resolution: 1W.
Setting Accuracy: ? (0.5% + 2 W).
Regulation: <2% for a 20W to 280W power change.
Slew Rate Range: <0.1W/ms to >3,200W/ms
LEVEL CONTROL
Control Modes:
Level A, Level B, Transient Internal (switching between A and B using the
built-in generator), Transient External (switching between A and B using an
external switching signal), and Remote (level proportional to an external
voltage).
Level A, Level B:
Each set by 10-turn potentiometer. Range, resolution and setting accuracy as
shown above.
Slew Rate Control
All changes of level are subjected to slew rate control which is adjustable over
a 30,000:1 range by a 3 position switch and vernier.
Input Enable
Latching switch which turns the load condition on.
Slow Start
Engages slew rate control to limit the speed of the ON transition when the input
is enabled.
Dropout Voltage
Defines a minimum voltage below which the load current will reduce to zero (as
may be required for testing batteries or power sources that can not be short
circuited). Adjustable from 0V to 80V.
TRANSIENT GENERATOR
Pulse Repetition Rate: 0.1Hz to 10kHz in 3 overlapping
ranges.
Setting Resolution: 0.1Hz, 1Hz, 10Hz depending on range.
Setting Accuracy: ? 2% of range.
Pulse Duty Cycle: 1% to 99% continuously variable.
Sync Output: Pulse synchronous with transient level change.
DISPLAYS and METERING
Dual 4 digit LED displays showing a variety of settings or
measurements as set by a five position switch.
Setting Displays: Level A and Level B, Frequency and Duty Cycle, Slew Rate and
Drop Out.
Measurements: Amps and Volts, Watts and Ohms (calculated from V and I
measurements).
Meter Accuracy: Voltage ?(0.1% + 1 digit), Current ?(0.2% + 20mA).
REMOTE VOLTAGE SENSE
Remote or Local voltage sensing can be selected via a rear panel switch. Sense
connections are on the rear panel terminal block.
Input Impedance: 680kW (each input) to load negative.
Max. Voltage Drop: 6 volts.
REMOTE CONTROL
Analogue - the level for any mode can be set by an external
analogue voltage (4V full scale).
TTL - the change between the Level A and Level B settings for any mode can be
done in response to an external logic signal.
The inputs for analogue or TTL have an impedance of 400k Ohms to the power
negative terminal and a common mode range of ?100V.
Remote Disable - when the front panel input enable switch is in the on position,
this input can be used to turn the load on or off in response to an external
logic signal. Fully opto isolated input.
CURRENT MONITOR
An analogue signal representing the current waveform is
available for connection to an oscilloscope or external meter. Scaling is
50mV/Amp and compliance is ?3V with respect to the load negative terminal.
PROTECTION
The load incorporates both a power limiter circuit and a
fault trip which responds to excess voltage, current, power or temperature. The
trip puts the load into a high impedance state. Surge suppessors are
incorporated to limit transient voltage spikes.
GENERAL
Rear Terminals: Input via 6mm terminal/binding posts. Remote
sense, current monitor, remote control via terminal block.
Front Terminals: Input via 4mm terminal/binding posts (30A max.). Current
monitor via 4mm sockets.
Power Requirement: 220V - 240V or 110V - 120V ?10%, 50/60Hz. Installation
Category II. Consumption 30VA max.
Operating Range: +5oC to +40oC, 20% to 80% RH.
Storage Range: -40oC to +70oC.
Environmental: Indoor use at altitudes up to 2000m. Pollution degree 2.
Safety and EMC: Complies with EN61010-1 and EN61326.
Size: 130 x 212 x 435mm [HxWxD], (3U x half rack).
Weight: 6kg.
Option:
19-inch rack mount kit.
|